User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2718 DATE: 24-Mar-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 90% NUMBER OF TESTS: 29 NUMBER OF LINES: 118 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 6060B ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ D N 1.002 ASK- R P J F W 1.003 HEAD {DC-VOLTAGE ACCURACY TEST} 1.004 DISP Four "C" size batteries must be installed. 1.004 DISP Verify the UUT backlit by putting the UUT in a dart 1.004 DISP area ( applicable on PM 2718/3XX ). 1.004 DISP Set the function switch to V-DC . 1.004 DISP Press key until UUT displays "000.0 mV *". 1.004 DISP Connect a banana cable to UUT OHM-Hz-V-mA (red terminal) 1.004 DISP and O (black terminal). Place a short across the end of 1.004 DISP the cable. Then press key on UUT. 1.005 DISP Remove the short and connect the cable to MULTIFUNCTION 1.005 DISP CALIBARTOR Output. 1.006 5100 1 100mV 0.24U O 2W 2.001 DISP Set UUT to 10V range by pressing key once. 2.002 5100 10 10V 0.006U 2W 3.001 5100 10 -10V 0.006U 2W 4.001 DISP Set UUT to 100V range by pressing key once. 4.002 5100 100 100V 0.06U 2W 5.001 DISP Set UUT to 1000V range by pressing key once. 5.001 DISP CAUTION : [27][91]7m HIGH VOLTAGE ! [27][91]m 5.002 5100 1000 900V 1.1U 2W 6.001 5100 0V S 2W 6.002 5100 * S 6.003 RSLT = 6.004 HEAD {AC-VOLTAGE ACCURACY TEST} 6.005 DISP Set the function switch to V-AC . 6.005 DISP Set UUT to 1V range by pressing key until UUT 6.005 DISP displays "~ 000.0 mV *". 6.006 5100 1 100mV 1.3U 50H 2W 7.001 5100 1 100mV 20U 50kH 2W 8.001 DISP Set UUT to 10V range by pressing key once. 8.002 5100 10 10V 0.04U 50H 2W 9.001 5100 10 10V 0.65U 50kH 2W 10.001 DISP Set UUT to 100V range by pressing key once. 10.002 5100 100 100V 0.4U 50H 2W 11.001 5100 100 100V 0.4U 1kH 2W 12.001 DISP Set UUT to 1000V range by pressing key once. 12.001 DISP CAUTION : [27][91]7m HIGH VOLTAGE ! [27][91]m 12.002 5100 1000 900V 3.7U 50H 2W 13.001 5100 1000 900V 3.7U 1kH 2W 14.001 5100 * S 14.002 RSLT = 14.003 HEAD {RESISTANCE ACCURACY TEST} 14.004 DISP Set the function switch to . 14.004 DISP Press the key until UUT displays " OL. *". 14.004 DISP Disconect the cable from MULTIFUNCTION CALIBRATOR 14.004 DISP Output and place a short across the cable. 14.004 DISP Press key then remove the short from the cable. 14.004 DISP Connect the cable to MULTIFUNCTION CALIBRATOR Output. 14.005 RESF 1000 1000Z 2U 2W 15.001 DISP Set UUT to 10 KOhm range by pressing key once. 15.002 RESF 10 10kZ 0.02U 2W 16.001 DISP Set UUT to 100 KOhm range by pressing key once. 16.002 RESF 100 100kZ 0.2U 2W 17.001 DISP Set UUt to 1 MOhm range by pressing key once. 17.002 RESF 1 1MZ 0.002U 2W 18.001 DISP Set UUT to 10 MOhm range by pressing key once. 18.002 RESF 10 10MZ 0.04U 2W 19.001 RESF * S 19.002 RSLT = 19.003 HEAD {CONTINUITY AND BEEPER TEST} 19.004 DISP Set the function switch to . 19.005 RESF 1Z S 2W 19.006 EVAL Does the UUT displays "00X.X" and activates the BEEPER ? 20.001 RESF * S 20.002 RSLT = 20.003 HEAD {FREQUENCY ACCURACY TEST} 20.004 DISP Set the function switch to . 20.004 DISP Disconnect the cable from MULTIFUNCTION CALIBRATOR 20.004 DISP Output and connect it to the 6060B RF Output. 20.005 6060 19kH 0.019U 1VP -D 21.001 6060 199kH 0.199U 1VP -D 22.001 6060 * S 22.002 RSLT = 22.003 HEAD {DC-CURRENT ACCURACY TEST} 22.004 DISP Set the function switch to . Press key until 22.004 DISP UUT displays "00.00 mA *". 22.004 DISP Move the cable to MULTIFUNCTION CALIBRATOR Output from 22.004 DISP the 6060 RF Output. 22.005 5100 20 20mA 0.1U 2W 23.001 5100 20 -20mA 0.1U 2W 24.001 DISP Set UUT to 200mA range by pressing key once. 24.002 5100 200 200mA 1U 2W 25.001 DISP Move the red terminal to <10A> input from V-Ohm input. 25.001 DISP Press key until UUT displays "0.000 A *". 25.001 DISP CAUTION : [27][91]7m HIGH CURRENT ! [27][91]m 25.002 5100 2 1A 0.006U 2W 26.001 5100 * S 26.002 RSLT = 26.003 HEAD {AC - CURRENT ACCURACY TEST} 26.004 DISP Set function switch to . 26.004 DISP Press key until UUT displays "~ 0.000 A *". 26.004 DISP CAUTION : [27][91]7m HIGH CURRENT ! [27][91]m 26.005 5100 2 1A 0.01U 50H 2W 27.001 5100 * S 27.002 DISP Move the red terminal to input from <10A> 27.002 DISP input. 27.002 DISP Press key until the UUT displays "~00.00 mA *". 27.003 5100 20 20mA 0.14U 50H 2W 28.001 DISP Set UUT to 200mA range by pressing once. 28.002 5100 200 200mA 1.4U 50H 2W 29.001 5100 * S 29.002 RSLT = 29.003 HEAD THE END